Abstract (EN):
A PVC membrane electrode, without inner reference solution, based on an ion association extraction system responding to ephedrine is described. It incorporates an ephedrine-tetrakis (4-chlorophenyl)borate ion-pair complex in 2-nitrophenyloctyl ether. The prepared electrode exhibits a near Nernstian response (57.5 mV per decade) over the concentration range of 2 x 10(-5) 10(-1) M ephedrine in solutions of pH 2.5-9. The reproducibility of the electrode potentials were +/-1 mV by day during at least 6 months. Response time was about 6 s for ephedrine concentrations between 10(-5) and 10(-1) M. Determinations of ephedrine in pharmaceutical preparations (tablets, nasal drops and syrups) by direct potentiometry gave an average recovery of 99.1% (w/w) and a mean standard deviation of 1.9% (w/w).
Idioma:
Inglês
Tipo (Avaliação Docente):
Científica
Nº de páginas:
5