Abstract (EN):
A 41-year-old man with history of surgical replacement of the aortic valve with a 21mm-Mitroflow bioprosthesis (1A), presented with functional class IV heart failure. Transesophageal echocardiography confirmed severe bioprosthesis obstruction (1B). We implanted a 23mm-CoreValve® Evolut¿ R (Medtronic, Minneapolis, USA) in the aortic bioprosthesis, by transfemoral approach. The valve was recaptured and repositioned during deployment (1C-E). Immediate (1F), one (1G) and three-month (1H) transthoracic echocardiography confirmed significant reduction in transaortic gradients. The patient remained in functional class I.
Language:
English
Type (Professor's evaluation):
Scientific
No. of pages:
1