Coupling of multi-pumping flow systems to flame atomic spectrometric detectors: Preliminary studies and analytical applications in the analysis of biological samples

Título
Coupling of multi-pumping flow systems to flame atomic spectrometric detectors: Preliminary studies and analytical applications in the analysis of biological samples
Tipo
Artigo em Revista Científica Internacional
Ano
2006
Autores
Lopes, C.M.P.V
(Autor)
Outra
A pessoa não pertence à instituição. A pessoa não pertence à instituição. A pessoa não pertence à instituição. Sem AUTHENTICUS Sem ORCID
Santos,J.L.M.
(Autor)
FFUP
Ver página pessoal Sem permissões para visualizar e-mail institucional Pesquisar Publicações do Participante Sem AUTHENTICUS Sem ORCID
Revista
Título: Atomic SpectroscopyImportada do Authenticus Pesquisar Publicações da Revista
Vol. 27 5
Páginas: 156-164
ISSN: 0195-5373
Classificação Científica
FOS: Ciências médicas e da saúde > Outras ciências médicas
CORDIS: Ciências da Saúde
Outras Informações
ID Authenticus: P-004-HC9
Resumo (PT): The so-called "multi-pumping flow systems" (MPFS), which rely on the utilization of solenoid-actuated micro-pumps (SAMPs) for introduction and propulsion of solutions, are a relatively recent development in the flow analysis field. They have already been coupled to several molecular spectrometric detectors with claimed advantages. However, the applicability and potentialities of their coupling to flame atomic spectrometric (FAS) detectors has never been studied. This work describes the study of coupling multi-pumping flow systems to flame atomic absorption spectrometers (FAAS) and flame (atomic emission) photometers (FP). The main purpose was to study the behavior of SAMPs when subjected to the aspiration effect inherent to nebulizers of FAS instrumentation (Venturi effect) and to evaluate the effects of the pulsed flow, a main characteristic of MPFS, over the efficiency of the nebulization and atomization processes. Based on these preliminary studies, several MPFS that enabled the on-line dilution of samples and/or the on-line addition of reagents (e.g., a releasing agent), two sample pre-treatment steps frequently required when using FAS detectors in analytical work, were developed. In this study, the analytical applicability has been demonstrated in the determination of Na in urine samples (using FP) and Ca and Mg in hair sample digests (using FAAS). The multi-pumping flow systems enable the configuration of simple analytical flow systems that are robust and capable of providing precise and accurate results with good sampling rates. <br> <br>
Abstract (EN): The so-called "multi-pumping flow systems" (MPFS), which rely on the utilization of solenoid-actuated micro-pumps (SAMPs) for introduction and propulsion of solutions, are a relatively recent development in the flow analysis field. They have already been coupled to several molecular spectrometric detectors with claimed advantages. However, the applicability and potentialities of their coupling to flame atomic spectrometric (FAS) detectors has never been studied. This work describes the study of coupling multi-pumping flow systems to flame atomic absorption spectrometers (FAAS) and flame (atomic emission) photometers (FP). The main purpose was to study the behavior of SAMPs when subjected to the aspiration effect inherent to nebulizers of FAS instrumentation (Venturi effect) and to evaluate the effects of the pulsed flow, a main characteristic of MPFS, over the efficiency of the nebulization and atomization processes. Based on these preliminary studies, several MPFS that enabled the on-line dilution of samples and/or the on-line addition of reagents (e.g., a releasing agent), two sample pre-treatment steps frequently required when using FAS detectors in analytical work, were developed. In this study, the analytical applicability has been demonstrated in the determination of Na in urine samples (using FP) and Ca and Mg in hair sample digests (using FAAS). The multi-pumping flow systems enable the configuration of simple analytical flow systems that are robust and capable of providing precise and accurate results with good sampling rates. <br> <br>
Idioma: Português
Tipo (Avaliação Docente): Científica
Documentos
Não foi encontrado nenhum documento associado à publicação.
Publicações Relacionadas

Da mesma revista

Sequential preconcentration for the ETAAS determination of Pd in human urine (2002)
Artigo em Revista Científica Internacional
Costa, RCC; Almeida, AA; Araujo, AN; Silva, VL
Quantification of Ag, Co, Si, and Zn in port wine by atomic absorption spectrometry (1995)
Artigo em Revista Científica Internacional
Soares, ME; Bastos, ML; Ferreira, MA

Ver todas (12)

Recomendar Página Voltar ao Topo
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Termos e Condições  I Acessibilidade  I Índice A-Z
Página gerada em: 2025-09-13 às 09:57:59 | Política de Privacidade | Política de Proteção de Dados Pessoais | Denúncias | Livro Amarelo Eletrónico