Saltar para:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Início > Publicações > Visualização > Validation of dose measurements by scintillating fiber optic dosimeters for medical applications

Publicações

Validation of dose measurements by scintillating fiber optic dosimeters for medical applications

Título
Validation of dose measurements by scintillating fiber optic dosimeters for medical applications
Tipo
Artigo em Livro de Atas de Conferência Internacional
Ano
2013
Autores
Correia, A
(Autor)
Outra
A pessoa não pertence à instituição. A pessoa não pertence à instituição. A pessoa não pertence à instituição. Sem AUTHENTICUS Sem ORCID
Pirraco, R
(Autor)
Outra
A pessoa não pertence à instituição. A pessoa não pertence à instituição. A pessoa não pertence à instituição. Sem AUTHENTICUS Sem ORCID
Rosa, CC
(Autor)
FCUP
Ata de Conferência Internacional
8th Iberoamerican Optics Meeting and 11th Latin American Meeting on Optics, Lasers, and Applications
Porto, PORTUGAL, JUL 22-26, 2013
Indexação
Publicação em ISI Web of Knowledge ISI Web of Knowledge - 0 Citações
Publicação em Scopus Scopus - 0 Citações
Classificação Científica
FOS: Ciências exactas e naturais > Física
Outras Informações
ID Authenticus: P-006-K44
Abstract (EN): Organic scintillators have been promoted and widely used in scintillating fiber-optic dosimeters (SFOD) due to their tissue-equivalent characteristics, small sensitive volume combined with high spatial resolution, and emission of visible light proportional to the absorbed electron and gamma dose rate. In this paper we will present the validation of Monte Carlo simulations of dose measurements assisted by scintillating fiber optic dosimeters operating in the visible spectral range, in the context of the development of fiber optic dosimeters targeted to Brachytherapy. The Monte Carlo simulation results are compared to measurements performed with SFOD test probes, assembled with BCF-60 (Saint Gobain) samples of 1 mm diameter and 0.35 to 1.5 cm length, coupled to PMMA optical fiber. The optical signal resulting from scintillation and Cherenkov light is transmitted through an additional optical fiber link to a remote measuring device. For SFOD probes irradiation a dedicated PMMA phantom was used. The results were validated against measurements obtained with a properly calibrated pinpoint ionization chamber (PTW). The probes were positioned in a radial arrangement, with a radioactive source at its center point. The gamma-rays source is a Nucletron Microselectron-V2 Ir-192. The dose curves are obtained according to the different positions in the phantom with the SFOD dosimeters. The system is able to use a Fiber Optic Multiplexer (FOM) controlled with Labview software.
Idioma: Inglês
Tipo (Avaliação Docente): Científica
Nº de páginas: 7
Documentos
Não foi encontrado nenhum documento associado à publicação.
Recomendar Página Voltar ao Topo
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Termos e Condições  I Acessibilidade  I Índice A-Z
Página gerada em: 2025-10-03 às 05:01:53 | Política de Privacidade | Política de Proteção de Dados Pessoais | Denúncias | Livro Amarelo Eletrónico