Abstract (EN):
A new methodology for the assessment of thiocyanate (SCN(-)) is proposed based on optical fiber (OF) detection coupled to a liquid chromatography system (LC). The developed methodology showed an adequate performance for the analysis of SCN(-) comparable to a high performance liquid chromatography with UV detector (HPLC-UV) methodology: a detection limit of 3 mu g L(-1), a linear range from 4 to 400 mu g L(-1), and an analytical time of less than 6 min. The OF based methodology was of compact design and easy operation. This simple system has the potential to be used as a sensing approach for SCN(-) in seawater.
Language:
English
Type (Professor's evaluation):
Scientific
Contact:
lisilva@ua.pt
No. of pages:
5