Abstract (EN):
The construction and evaluation of a quinidine ion-selective electrode, without an inner reference solution, based on an ion exchanger [quinidine tetrakis(4-chlorophenyl)borate] dissolved in 2-nitrophenyl octyl ether and immobilized in PVC is described. The prepared electrode showed a linear response in the concentration range 3 x 10(-5) - 1 x 10(-2) M in quinidine, with a slope of 57.5 mV per decade when the pH of the solutions was between 5 and 7.5. The reproducibility of the electrode was approximately +/- 1 mV during a day's work, with a response time of less than 10 s for quinidine concentrations between 10(-4) and 10(-2) M. The electrode was used for the direct potentiometric determination of quinidine in pharmaceutical preparations, giving results with an average recovery of 100.7% and a mean standard deviation of 2.3%.
Language:
English
Type (Professor's evaluation):
Scientific
No. of pages:
5