Go to:
Logótipo
Você está em: Start > Publications > View > Age-dependent calibration factors for in-vivo monitoring of I-131 in thyroid using Monte Carlo simulations
Map of Premises
Principal
Publication

Age-dependent calibration factors for in-vivo monitoring of I-131 in thyroid using Monte Carlo simulations

Title
Age-dependent calibration factors for in-vivo monitoring of I-131 in thyroid using Monte Carlo simulations
Type
Article in International Scientific Journal
Year
2019
Authors
Gomez Ros, JM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Moraleda, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teles, P
(Author)
Other
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Tyminska, K
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Saizu, MA
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Gregoratto, D
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Lombardo, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Berkovsky, V
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Ratia, G
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Broggio, D
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
Vol. 125
Pages: 96-105
ISSN: 1350-4487
Publisher: Elsevier
Other information
Authenticus ID: P-00Q-JW7
Abstract (EN): Following a nuclear incident, I-131 monitoring is usually carried out by performing in vivo measurements of the retained activity in the thyroid. The measurement equipment needs to be calibrated in order to convert thyroid count rates into retained activity. This calibration relies on measurements or Monte Carlo computations using appropriate anthropomorphic model of the thyroid gland contaminated with a known activity of the radionuclide of interest. Several factors influence the calibration procedure, such as the geometry, the individual anatomical characteristics, and the presence of I-131 in other organs and tissues of the body. In particular, given their age-dependent anatomical changes, appropriate values of detector efficiencies for children are necessary to reduce results uncertainties when children are involved. In this study, Monte Carlo simulations for several detectors and phantoms of different ages have been performed to investigate the variation of the detection efficiency with distance, age and thyroid volume. It was found that the counting efficiency varies linearly with the thyroid volume and inversely with the squared detector-phantom distance. Age-dependent correction factors to be applied to the adult calibration factor have been derived and can be used when calibration factors for children are not available.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 10
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

TLD measurements and Monte Carlo simulations for glandular dose and scatter fraction assessment in mammography: A comparative study (2011)
Article in International Scientific Journal
Di Maria, S; Barros, S; Bento, J; Teles, P; Figueira, C; Pereira, M; Vaz, P; Paulo, G
Performance parameters of a whole body counter (2010)
Article in International Scientific Journal
Bento, J; Teles, P; Silva, L; Nogueira, P; Neves, M; Vaz, P
Monte Carlo modelling for the in vivo lung monitoring of enriched uranium: Results of an international comparison (2012)
Article in International Scientific Journal
Broggio, D; Bento, J; Caldeira, M; Cardenas Mendez, E; Farah, J; Fonseca, T; Konvalinka, C; Liu, L; Perez, B; Capello, K; Cowan, P; Cruzate, JA; Freire, L; Gomez Ros, JM; Gossio, S; Heide, B; Huikari, J; Hunt, J; Kinase, S; Kramer, GH...(mais 21 authors)
Child and adult thyroid monitoring after a reactor accident (CAThyMARA): Technical recommendations and remaining gaps (2019)
Article in International Scientific Journal
Broggio, D; Baude, S; Belchior, A; Berkovskyy, V; Bonchuck, Y; Dewoghelaere, J; Etherington, G; Fojtik, P; Franck, D; Gomez Ros, JM; Gregoratto, D; Helebrant, J; Dubreuil, GH; Hulka, J; Isaksson, M; Kocsonya, A; Lebacq, AL; Likhtarev, I; Lombardo, P; Lopez, MA...(mais 22 authors)
Assessment from in vivo measurements of thyroid dose due to iodine-131 inhalation when stable iodine has been administered (2019)
Article in International Scientific Journal
Broggio, D; Teles, P; Vrba, T; Berkovskyy, V

See all (6)