Abstract (EN):
Background: The purpose of this study was to evaluate the
morphological alterations of the ProTaper next rotary file (PTN)
under scanning electron microscopy (SEM).
Materials and Methods: A total of 18 simulated root canals were
allocated to three groups. Six new sets of PTN instruments were used
3 times. A #10 K-file was inserted into the working length, followed
by ProGlider to create a glide path. Ensuring the manufacturer¿s
instructions with 99% ethyl alcohol for irrigation, all canals were
prepared. Files were photographed in the same position before and
after three canals preparations using a high-resolution SEM.
Result: A metal strip appeared on one X1 instrument surface
preoperatively. Microcrack defects were observed on two X2 files
postoperatively, and the blunt cutting edge was observed on three
X1 files before and after use and one file fractured.
Conclusion: Small number of changes appeared on PTN surfaces,
yet same PTN file can be used safely 3 times to prepare multi-rooted
teeth within the same patient.
Language:
English
Type (Professor's evaluation):
Scientific
No. of pages:
6
License type: