Abstract (EN):
In this work, a hybrid interferometer for simultaneous measurement of the partial pressures of O-2 and CO2 mixtures is reported. The sensing head consist in two different interferometers based on a Fabry-Perot cavity and a modal interference configuration. The intrinsic FP cavity was created by splicing a single mode fiber (SMF28) with a graded index fiber section that was then subjected to chemical etching creating a cavity. The second interferometer is based on a splice of a pure silica tube in series with the Fabry-Perot. Due to the design, different sensitivities are achieved for the pressure inducing refractive index changes of each gas. The rms deviations were found to be +/- 0.079 kPa and +/- 0.029 kPa for CO2 and O-2 partial pressure measurements, respectively.
Language:
English
Type (Professor's evaluation):
Scientific