Abstract (EN):
A procedure is described for the colorimetric determination of iron in infant fortified formulas based on sequential injection analysis (SIA). Iron(III) complexation with thiocyanate is used as colour developing reaction. The system enables the determination of iron in the samples (after digestion by dry ashing and treatment with 0.2 mol/L nitric acid in the range of 0.50-20.0 mg/L, consuming 140 mu L of the sample and 8 mg thiocyanate per determination. The reactor geometry and the adjustment of the ionic content of the calibration solutions is important for the accuracy of the results. A regression line according to the equation [Fe(III) (mg/L)](SIA) = - 0.3(+/- 0.4) + 1.03(+/- 0.04) [Fe(III) (mg/L)](FAAS) was obtained after comparative analysis of a set of 12 samples. The measurement rate was 34 s, thus allowing to analyze 100 samples per hour with a relative standard deviation lower than 2%.