Go to:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Start > Publications > View > Self-calibrating total-mass controller for the neuromuscular blockade matching the anesthesiologists' mindset
Publication

Publications

Self-calibrating total-mass controller for the neuromuscular blockade matching the anesthesiologists' mindset

Title
Self-calibrating total-mass controller for the neuromuscular blockade matching the anesthesiologists' mindset
Type
Article in International Conference Proceedings Book
Year
2013
Authors
Luis A Paz
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Margarida M Silva
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Simao Esteves
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teresa Mendonca
(Author)
FCUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page Without ORCID
Conference proceedings International
Pages: 723-728
21st Mediterranean Conference on Control and Automation (MED)
Platanias, GREECE, JUN 25-28, 2013
Scientific classification
FOS: Engineering and technology > Electrical engineering, Electronic engineering, Information engineering
Other information
Authenticus ID: P-008-EXY
Abstract (EN): A self-calibrating total-mass controller for the neuromuscular blockade (NMB) based on the minimally parameterized parsimonious (MPP) Wiener model is presented. Using input-output data collected from the initial bolus administration until recovery, the parameters of the MPP model that are used for the controller design are chosen from a set of fifty realistic models for the effect of the muscle relaxant rocuronium in the NMB. In order to overcome modeling uncertainties a numeric self-calibration of the parameters in the MPP model nonlinearity was implemented. This step matches the anesthesiologists' mindset. The controller was extensively tested in simulation and in real cases. The good reference tracking results show the reliability of the proposed strategy and supports its use in the clinical practice.
Language: English
Type (Professor's evaluation): Scientific
Contact: lpaz@fc.up.pt; margarida@fc.up.pt; silva@fc.up.pt; tmendo@fc.up.pt
No. of pages: 6
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same authors

Automated Total IntraVenous Anesthesia (amTIVA) from induction to recovery (2014)