Abstract (EN):
This paper reports the development of a novel electrochemical assay for xanthohumol (XN) by square-wave adsorptive-stripping voltammetry (SWAdSV) with a hanging mercury drop electrode. The method showed good repeatability (CV < 2%) and linearity (between 10 and 250 mu g L-1), as well as suitable limits of detection (2.6 mu g L-1) and quantification (8.8 mu g L-1). The method was applied for the quantification of this compound in spent hops, and the results obtained were compared with the HPLC-UV method. XN contents determined by the SWAdSV method were 16 +/- 1 and 100 +/- 4 mu g L-1 for aqueous and methanolic extracts, respectively. The developed new methodology considerably reduces the analysis time, approximately from 25 min (HPLC-UV method) to 7 min, enabling a high sample throughput. In addition, the detection and quantification limits were approximately 5-fold lower than those obtained with the chromatographic method.
Language:
English
Type (Professor's evaluation):
Scientific
Contact:
lfguido@fc.up.pt
No. of pages:
5