Design criteria in choosing optimized OCT scanning regimes - art. no. 66271E

Title
Design criteria in choosing optimized OCT scanning regimes - art. no. 66271E
Type
Article in International Conference Proceedings Book
Year
2007
Authors
Carla Carmelo Rosa
(Author)
FCUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
John Rogers
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Justin Pedro
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Richard Rosen
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Adrian Podoleanu
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Conference proceedings International
Pages: E6271-E6271
Conference on Optical Coherence Tomography and Coherence Techniques III
Munich, GERMANY, JUN 00-19, 2007
Indexing
Publicação em ISI Web of Knowledge ISI Web of Knowledge - 0 Citations
Publicação em Scopus Scopus - 0 Citations
Other information
Authenticus ID: P-004-E4B
Abstract (EN): A comparative analysis on the performance of different scanning regimes in time domain optical coherence tomography is presented in terms of image size. Safety thresholds due to the different continuous irradiation time per transverse pixel in different scanning regimes are also considered. We present the maximum exposure level for a variety of scanning procedures, employing either A scanning (depth priority) or T scanning (transverse priority) when generating cross section images, en face images or collecting 3D volumes. We present a comparison between such B-scan images, and different criteria to allow the user to choose the right mode of operation. Mainly, two criteria are detailed, a scanning criterion and a safety criterion. The scanning criterion depends on the number of pixels along the lateral and axial directions. The analysis shows that en-face scanning allows wider images while the longitudinal scanning is more suitable to deep cross sections. The safety criterion refers to safety levels to be observed in each scanning mode. We show that the flying spot OCT imaging has different safety limits for T- and A- based imaging modes. The analysis leads to maximum permissible optical power levels that favors T-scan imaging of wide objects. We then apply the analysis considering as object the eye.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 9
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same authors

Multiscan time-domain optical coherence tomography for retina imaging (2007)
Article in International Scientific Journal
Carla Carmelo Rosa; John Rogers; Justin Pedro; Richard Rosen; Adrian Podoleanu
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-07-15 at 11:18:07 | Privacy Policy | Personal Data Protection Policy | Whistleblowing