FATIGUE TEST ON A 3-ELEMENT IMPLANT BRIDGE COMBINING A CONVENTIONAL AND A DYNAMIC ABUTMENT

Title
FATIGUE TEST ON A 3-ELEMENT IMPLANT BRIDGE COMBINING A CONVENTIONAL AND A DYNAMIC ABUTMENT
Type
Article in International Conference Proceedings Book
Year
2015
Authors
Ferreira, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Oliveira, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Song, NN
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Machado, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Conference proceedings International
Pages: 1481-1488
6th International Conference on Mechanics and Materials in Design
Ponta Delgada/Azores, 26-30 July 2015
Indexing
Publicação em ISI Web of Knowledge ISI Web of Knowledge - 0 Citations
Other information
Authenticus ID: P-00K-MYV
Abstract (EN): This work compares the fatigue performance of a conventional abutment with a dynamic abutment, both part of a 3-element implant fixed bridge. To perform this test, the ISO standard 14801 was used. A compressive static test was performed to define the initial reference load (3005N). The samples were than subjected to progressively decreasing loads until a total of five million cycles were reached. At that point (1050N), three samples were subjected to the same load obtaining the same result. The comparison of these results with the values for mastication loads found on bibliography, highlight the safety of use of the dynamic abutments.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 8
Documents
We could not find any documents associated to the publication.
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-08-24 at 13:10:24 | Privacy Policy | Personal Data Protection Policy | Whistleblowing