Abstract (EN):
Background and objectives: According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods: Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results: The difference between the predicted and the real BIS was in average 30.09 +/- 18.73 s. Conclusion: Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications. (C) 2019 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda.