Go to:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Start > Publications > View > Calculated beam quality correction factors for ionization chambers in MV photon beams
Publication

Publications

Calculated beam quality correction factors for ionization chambers in MV photon beams

Title
Calculated beam quality correction factors for ionization chambers in MV photon beams
Type
Article in International Scientific Journal
Year
2020
Authors
Tikkanen, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Zink, K
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Pimpinella, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teles, P
(Author)
FCUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Borbinha, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Ojala, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Siiskonen, T
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Goma, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Pinto, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
Vol. 65
ISSN: 0031-9155
Other information
Authenticus ID: P-00R-Y4P
Abstract (EN): The beam quality correction factor, , which corrects for the difference in the ionization chamber response between the reference and clinical beam quality, is an integral part of radiation therapy dosimetry. The uncertainty of is one of the most significant sources of uncertainty in the dose determination. To improve the accuracy of available data, four partners calculated factors for 10 ionization chamber models in linear accelerator beams with accelerator voltages ranging from 6 MV to 25 MV, including flattening-filter-free (FFF) beams. The software used in the calculations were EGSnrc and PENELOPE, and the ICRU report 90 cross section data for water and graphite were included in the simulations. Volume averaging correction factors were calculated to correct for the dose averaging in the chamber cavities. A comparison calculation between partners showed a good agreement, as did comparison with literature. The values from TRS-398 were higher than our values for each chamber where data was available. The values for the FFF beams did not follow the same , relation as beams with flattening filter (values for 10 MV FFF beams were below fits made to other data on average by 0.3%), although our FFF sources were only for Varian linacs.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 15
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

Monte Carlo calculation of quality correction factors based on air kerma and absorbed dose to water in medium energy x-ray beams (2020)
Article in International Scientific Journal
Czarnecki, D; Zink, K; Pimpinella, M; Borbinha, J; Teles, P; Pinto, M
Medical staff extremity dosimetry in CT fluoroscopy: an anthropomorphic hand voxel phantom study (2013)
Article in International Scientific Journal
Figueira, C; Becker, F; Blunck, C; DiMaria, S; Baptista, M; Esteves, B; Paulo, G; Santos, J; Teles, P; Vaz, P
Determination of consensus k(Q) values for megavoltage photon beams for the update of IAEA TRS-398 (2020)
Article in International Scientific Journal
Andreo, P; Burns, DT; Kapsch, RP; McEwen, M; Vatnitsky, S; Andersen, CE; Ballester, F; Borbinha, J; Delaunay, F; Francescon, P; Hanlon, MD; Mirzakhanian, L; Muir, B; Ojala, J; Oliver, CP; Pimpinella, M; Pinto, M; de Prez, LA; Seuntjens, J; Sommier, L...(mais 4 authors)
A System for Digital Processing of Ophthalmic Images (1994)
Article in International Scientific Journal
Ana Maria Mendonça; Aurélio Campilho; José R. Nunes
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-07-15 at 11:06:11 | Privacy Policy | Personal Data Protection Policy | Whistleblowing