Go to:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Start > Publications > View > Monte Carlo simulation of the movement and detection efficiency of a whole-body counting system using a BOMAB phantom
Publication

Publications

Monte Carlo simulation of the movement and detection efficiency of a whole-body counting system using a BOMAB phantom

Title
Monte Carlo simulation of the movement and detection efficiency of a whole-body counting system using a BOMAB phantom
Type
Article in International Scientific Journal
Year
2012
Authors
Bento, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Barros, S
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teles, P
(Author)
Other
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Neves, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Goncalves, I
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Corisco, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Vaz, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. View Authenticus page Without ORCID
Journal
Vol. 148
Pages: 403-413
ISSN: 0144-8420
Other information
Authenticus ID: P-002-CP6
Abstract (EN): This study reports on the computational analysis and experimental calibration of the whole-body counting detection equipment at the Nuclear and Technological Institute (ITN) in Portugal. Two state-of-the-art Monte Carlo simulation programmes were used for this purpose: PENELOPE and MCNPX. This computational work was undertaken as part of a new set of experimental calibrations, which improved the quality standards of this studys WBC system. In these calibrations, a BOMAB phantom, one of the industry standards phantoms for WBC calibrations in internal dosimetry applications, was used. Both the BOMAB phantom and the detection system were accurately implemented in the Monte Carlo codes. The whole-body counter at ITN possesses a moving detector system, which poses a challenge for Monte Carlo simulations, as most codes only accept static configurations. The continuous detector movement was approximately described in the simulations by averaging several discrete positions of the detector throughout the movement. The computational efficiency values obtained with the two Monte Carlos codes have deviations of less than 3.2 , and the obtained deviations between experimental and computational efficiencies are less than 5 . This work contributes to demonstrate the great effectiveness of using computational tools for understanding the calibration of radiation detection systems used for in vivo monitoring.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 11
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

A REVIEW OF RADIATION PROTECTION REQUIREMENTS AND DOSE ESTIMATION FOR STAFF AND PATIENTS IN CT FLUOROSCOPY (2017)
Another Publication in an International Scientific Journal
Teles, P; Nikodemova, D; Bakhanova, E; Becker, F; Knezevic, Z; Pereira, MF; Sarmento, S
Study of nuclear medicine practices in Portugal from an internal dosimetry perspective (2012)
Article in International Scientific Journal
Bento, J; Teles, P; Neves, M; Santos, AI; Cardoso, G; Barreto, A; Alves, F; Guerreiro, C; Rodrigues, A; Santos, JAM; Capelo, C; Parafita, R; Martins, B
PERFORMANCE ASSESSMENT AND UNCERTAINTY EVALUATION OF A PORTABLE NAI-BASED DETECTION SYSTEM USED FOR THYROID MONITORING (2012)
Article in International Scientific Journal
Bento, J; Martins, B; Teles, P; Neves, M; Colarinha, P; Alves, F; Teixeira, N; Vaz, P; Zankl, M
On the Placement of Apron Dosemeters and Dose Assessment in Interventional Cardiology Procedures: Preliminary Results (2022)
Article in International Scientific Journal
Ferrari, P; Becker, F; Campani, L; Jansen, J; Jovanovie, Z; Krstie, D; Mariotti, F; Teles, P; Venturi, G
Human exposure to indoor radon: a survey in the region of Guarda, Portugal (2013)
Article in International Scientific Journal
Louro, A; Peralta, L; Soares, S; Pereira, A; Cunha, G; Belchior, A; Ferreira, L; Gil, OM; Louro, H; Pinto, P; Rodrigues, AS; Silva, MJ; Teles, P

See all (14)

Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-07-15 at 21:23:34 | Privacy Policy | Personal Data Protection Policy | Whistleblowing