Medical staff extremity dosimetry in CT fluoroscopy: an anthropomorphic hand voxel phantom study

Title
Medical staff extremity dosimetry in CT fluoroscopy: an anthropomorphic hand voxel phantom study
Type
Article in International Scientific Journal
Year
2013
Authors
Figueira, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Becker, F
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Blunck, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
DiMaria, S
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Baptista, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Esteves, B
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Paulo, G
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Santos, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teles, P
(Author)
Other
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Vaz, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. View Authenticus page Without ORCID
Journal
Vol. 58
Pages: 5433-5448
ISSN: 0031-9155
Other information
Authenticus ID: P-006-862
Abstract (EN): This work aims to contribute to the study of the radiation dose distribution delivered to the hands of medical staff members during a general computed tomographic (CT) fluoroscopic guided procedure. In this study, both Monte Carlo simulations and measurements were performed. For free-in-air and computed tomography dose index (CTDI) body phantom measurements, a standard pencil ionization chamber (IC) 100 mm long was used. The CT scanner model was implemented using MCNPX (Monte Carlo N-Particle eXtended) and was successfully validated by comparing the simulated results with measurements. Subsequently, CT images of a hand, together with an anthropomorphic phantom, were voxelized and used with the MCNPX code for dose calculations. The hand dose distribution study was performed both by using thermo-luminescent detector measurements and Monte Carlo simulations. The validated simulation tool provides a new perspective for detailed investigations of CT-irradiation scenarios. Simulations show that there is a strong dose gradient, namely the even zones of the hand that are in precise vicinity to the x-ray beam only receive about 4% of the maximum dose delivered to adjacent areas which are directly exposed to the primary x-ray beam. Finally, the scatter contribution of the patient was also studied through MC simulations. The results show that for directly exposed parts of the hand surface, the dose is reduced by the body of the patient (due to the shielding), whereas the dose is increased by scattered radiation from the patient for parts of the skin that receive scattered radiation only.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 16
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

Monte Carlo calculation of quality correction factors based on air kerma and absorbed dose to water in medium energy x-ray beams (2020)
Article in International Scientific Journal
Czarnecki, D; Zink, K; Pimpinella, M; Borbinha, J; Teles, P; Pinto, M
Determination of consensus k(Q) values for megavoltage photon beams for the update of IAEA TRS-398 (2020)
Article in International Scientific Journal
Andreo, P; Burns, DT; Kapsch, RP; McEwen, M; Vatnitsky, S; Andersen, CE; Ballester, F; Borbinha, J; Delaunay, F; Francescon, P; Hanlon, MD; Mirzakhanian, L; Muir, B; Ojala, J; Oliver, CP; Pimpinella, M; Pinto, M; de Prez, LA; Seuntjens, J; Sommier, L...(mais 4 authors)
Calculated beam quality correction factors for ionization chambers in MV photon beams (2020)
Article in International Scientific Journal
Tikkanen, J; Zink, K; Pimpinella, M; Teles, P; Borbinha, J; Ojala, J; Siiskonen, T; Goma, C; Pinto, M
A System for Digital Processing of Ophthalmic Images (1994)
Article in International Scientific Journal
Ana Maria Mendonça; Aurélio Campilho; José R. Nunes
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-07-14 at 19:45:27 | Privacy Policy | Personal Data Protection Policy | Whistleblowing