TLD measurements and Monte Carlo simulations for glandular dose and scatter fraction assessment in mammography: A comparative study

Title
TLD measurements and Monte Carlo simulations for glandular dose and scatter fraction assessment in mammography: A comparative study
Type
Article in International Scientific Journal
Year
2011
Authors
Di Maria, S
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Barros, S
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Bento, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Teles, P
(Author)
Other
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Figueira, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Pereira, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Vaz, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. View Authenticus page Without ORCID
Paulo, G
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
Vol. 46
Pages: 1103-1108
ISSN: 1350-4487
Publisher: Elsevier
Other information
Authenticus ID: P-002-KTR
Abstract (EN): The main purpose of this study was to validate and compare Mean Glandular Dose (MGD) values obtained using Monte Carlo simulations with experimental values obtained from Entrance Surface Dose (ESD) and depth dose measurements performed in a Hospital mammography unit. ESD and depth dose were measured using ThermoLuminescent Dosimeters (TLDs), and a tissue equivalent mammography phantom recommended by the American College of Radiology (ACR). Measurements and Monte Carlo simulations were also compared with the MGD calculated using the Automatic Exposure Control (AEC) system of the mammographic unit. In the simulations the Doppler energy broadening effect was also taken into account. The simulated ESD are about 5%-10% higher than the measured ESD values. The deviation between the measured and simulated MGD values in the phantom is of about 15%. The MGD evaluated using the AEC system is smaller both with respect to the Monte Carlo simulation and experimental result by a factor of about 15% and 25% respectively. Moreover the BackScatter Factor (BSF) estimated by Monte Carlo simulations was used for the MGD calculation according to the Wu's method. Finally the inclusion of the energy broadening effect on MGD calculation produces negligible variations on the simulated results.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 6
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

Performance parameters of a whole body counter (2010)
Article in International Scientific Journal
Bento, J; Teles, P; Silva, L; Nogueira, P; Neves, M; Vaz, P
Monte Carlo modelling for the in vivo lung monitoring of enriched uranium: Results of an international comparison (2012)
Article in International Scientific Journal
Broggio, D; Bento, J; Caldeira, M; Cardenas Mendez, E; Farah, J; Fonseca, T; Konvalinka, C; Liu, L; Perez, B; Capello, K; Cowan, P; Cruzate, JA; Freire, L; Gomez Ros, JM; Gossio, S; Heide, B; Huikari, J; Hunt, J; Kinase, S; Kramer, GH...(mais 21 authors)
Child and adult thyroid monitoring after a reactor accident (CAThyMARA): Technical recommendations and remaining gaps (2019)
Article in International Scientific Journal
Broggio, D; Baude, S; Belchior, A; Berkovskyy, V; Bonchuck, Y; Dewoghelaere, J; Etherington, G; Fojtik, P; Franck, D; Gomez Ros, JM; Gregoratto, D; Helebrant, J; Dubreuil, GH; Hulka, J; Isaksson, M; Kocsonya, A; Lebacq, AL; Likhtarev, I; Lombardo, P; Lopez, MA...(mais 22 authors)
Assessment from in vivo measurements of thyroid dose due to iodine-131 inhalation when stable iodine has been administered (2019)
Article in International Scientific Journal
Broggio, D; Teles, P; Vrba, T; Berkovskyy, V
Age-dependent calibration factors for in-vivo monitoring of I-131 in thyroid using Monte Carlo simulations (2019)
Article in International Scientific Journal
Gomez Ros, JM; Moraleda, M; Teles, P; Tyminska, K; Saizu, MA; Gregoratto, D; Lombardo, P; Berkovsky, V; Ratia, G; Broggio, D