TOF-Watch NMB Monitoring Misleading Display Output During Moderate Neuromuscular Blockade

Title
TOF-Watch NMB Monitoring Misleading Display Output During Moderate Neuromuscular Blockade
Type
Article in International Conference Proceedings Book
Year
2020
Authors
Couto, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Nunes, CS
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Amorim, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Conference proceedings International
Pages: 768-775
15th Mediterranean Conference on Medical and Biological Engineering and Computing (MEDICON)
UNESCO World Heritage Univ, Coimbra, PORTUGAL, SEP 26-28, 2019
Indexing
Publicação em ISI Web of Knowledge ISI Web of Knowledge - 0 Citations
Publicação em Scopus Scopus - 0 Citations
Other information
Authenticus ID: P-00R-F2N
Abstract (EN): In general anesthesia, Neuromuscular Blockade (NMB) agents are administered to ensure intubation quality and complete immobilization. Monitoring NMB, based on the evoked response evaluation after a peripheral nerve stimulation, is essential to provide insight on medication dosing and suitable approaches for NMB reversal. Professionals often rely on the data displayed on the monitor screen after TOF stimuli, assuming measurements present the expected fade and a TOF-ratio > 90% is enough to rule out residual paralysis. This can be inaccurate and mislead the clinician to extubate before adequate NMB curarization. Data from 31 patients that underwent general anesthesia were retrospectively analyzed. All patients received a standard dose of rocuronium (0.6 mg/kg) for intubation, during maintenance additional rocuronium boluses were administered if solicited. NMB monitoring was done continuously applying TOF stimuli at the adductor pollicis with TOF-Watch SX (R). Two types of monitoring display errors were studied: (1) valid TOF value without fade effect (invalid T1 > T2 > T3 > T4); (2) in the 30 min before extubation a valid TOF-ratio value > 90% (without error type 1) with T1 < 70%. Results show a mean (SD) of type 1 error on 42.0 (17.5)% of valid TOF measurements, type 2 error in 63 (45)%. Only 9 patients (29%) presented a TOF with no errors before extubation. These results may increase the distrust in the NMB monitoring devices, however its use should not be rejected when NMB agents are administered. Careful evaluation of the NMB is recommended and additional efforts should be placed in the accuracy of monitors data display to avoid errors.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 8
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same authors

Comparison of two pharmacokinetic-pharmacodynamic models of rocuronium bromide during profound neuromuscular block: analysis of estimated and measured post-tetanic count effect (2022)
Article in International Scientific Journal
Couto, M; Vide, S; Marco Arino, N; Troconiz, IF; Esteves, S; Nunes, CS; Amorim, P; Joaquim Mendes
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-08-11 at 10:39:47 | Privacy Policy | Personal Data Protection Policy | Whistleblowing