Go to:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Start > Publications > View > Miniaturized voltammetric cell for cathodic voltammetry making use of an agar membrane
Publication

Publications

Miniaturized voltammetric cell for cathodic voltammetry making use of an agar membrane

Title
Miniaturized voltammetric cell for cathodic voltammetry making use of an agar membrane
Type
Article in International Scientific Journal
Year
2018
Authors
Skalova, S
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Goncalves, LM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Navratil, T
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Barek, J
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Rodrigues, JA
(Author)
FCUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Vyskocil, V
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
The Journal is awaiting validation by the Administrative Services.
Vol. 821
Pages: 47-52
ISSN: 0022-0728
Other information
Authenticus ID: P-00P-079
Abstract (EN): Analysis of samples available in small volumes is an important challenge in analytical chemistry. Therefore, we have dealt with the development of a new miniaturized micro-volume voltammetric cell (MVVC) for the determination of reducible organic compounds in micro-volumes (20-50 mu L). The MVVC is composed of two compartments, a micro-volume compartment, where the analyte is present along with the working electrode, separated by means of an agar membrane (about 2 mm) from a large-volume compartment containing the supporting electrolyte and the counter and reference electrodes. This MVVC was tested by the determination of sodium anthraquione-2-sulfonate as a model substance for structurally similar anthraquinone-based anti-cancer drugs. Using differential pulse voltammetry at a polished silver solid amalgam electrode, the following analytical parameters were obtained: linear dynamic range from 1 mu mol L-1 to 1 mmol L-1, repeatability of 0.7% (on the concentration level 0.1 mmol L-1), and limit of detection of 0.15 mu mol L-1.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 6
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

MICROPIPETTE AS A TOOL FOR THE DETERMINATION OF THE IONIC SPECIES LIMITING THE POTENTIAL WINDOW AT LIQUID LIQUID INTERFACES (1991)
Another Publication in an International Scientific Journal
STEWART, AA; SHAO, Y; PEREIRA, CM; GIRAULT, HH
Spectroelectrochemical study of the copper(II) transfer assisted by 6,7-dimethyl-2,3-di(2-pyridyl)quinoxaline at the water vertical bar 1,2-dichloroethane interface (1998)
Article in International Scientific Journal
Tomaszewski, L; Ding, ZF; Fermin, DJ; Cacote, HM; Pereira, CM; Silva, F; Girault, HH
Passive direct methanol fuel cells acting as fully autonomous electrochemical biosensors: Application to sarcosine detection (2022)
Article in International Scientific Journal
Ferreira, NS; Carneiro, LPT; Viezzer, C; Almeida, MJT; Marques, AC; A.M.F.R. Pinto; Fortunato, E; Sales, MGF
Oxidation of ferrocene derivatives at a poly[Ni(saltMe)] modified electrode (2002)
Article in International Scientific Journal
Vilas Boas, M; Pereira, EM; Freire, C; Hillman, AR
MICRO-HOLE INTERFACE FOR THE AMPEROMETRIC DETERMINATION OF IONIC SPECIES IN AQUEOUS-SOLUTIONS (1994)
Article in International Scientific Journal
OSBORNE, MC; SHAO, Y; PEREIRA, CM; GIRAULT, HH

See all (12)

Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-08-06 at 12:09:50 | Privacy Policy | Personal Data Protection Policy | Whistleblowing