Go to:
Logótipo
Comuta visibilidade da coluna esquerda
Você está em: Start > Publications > View > Corrected and Improved Model for Educational Simulation of Neonatal Cardiovascular Pathophysiology
Publication

Publications

Corrected and Improved Model for Educational Simulation of Neonatal Cardiovascular Pathophysiology

Title
Corrected and Improved Model for Educational Simulation of Neonatal Cardiovascular Pathophysiology
Type
Another Publication in an International Scientific Journal
Year
2009
Authors
Zijlmans, M
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
van Meurs, WL
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Goodwin, JA
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Andriessen, P
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
Vol. 4
Pages: 49-53
ISSN: 1559-2332
Other information
Authenticus ID: P-003-N30
Abstract (EN): We identified errors in the software implementation of the mathematical model presented in: Sa Couto CD, van Meurs WL, Goodwin JA, Andriessen P. A model for educational simulation of neonatal cardiovascular pathophysiology. Simul Healthcare 2006; 1:4-12. Simulation results obtained with corrected code are presented for future reference. All but one of the simulation results do not differ by more than 9% from the previously published results. The heart rate response to acute loss of 30% of blood volume, simulated with corrected code is stronger than published target data. This modeling error was masked by errors in code implementation. We improved this response and the model by adjusting the gains and adding thresholds and saturations in the baroreflex model. General considerations on identification of model and code errors and model validity are presented. (Sim Healthcare 4:49-53, 2009)
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 5
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

¿Stan Vintage¿: A baseline patient for the Human Patient Simulator¿ with hemodynamic parameters from the scientific literature. (2006)
Article in International Scientific Journal
van Meurs, WL; Neto, P; Azevedo, H; Sá-Couto CD
Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z