On-line automated evaluation of lipid nanoparticles transdermal permeation using Franz diffusion cell and low-pressure chromatography

Title
On-line automated evaluation of lipid nanoparticles transdermal permeation using Franz diffusion cell and low-pressure chromatography
Type
Article in International Scientific Journal
Year
2016
Authors
Alves, AC
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Ramos, II
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Nunes, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Magalhaes, LM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Sklenarova, H
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Marcela A Segundo
(Author)
FFUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page Without ORCID
Reis, S
(Author)
FFUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Journal
Title: TalantaImported from Authenticus Search for Journal Publications
Vol. 146
Pages: 369-374
ISSN: 0039-9140
Publisher: Elsevier
Other information
Authenticus ID: P-00G-MAV
Abstract (EN): A low-pressure liquid chromatography system for the on-line quantification of caffeine loaded into lipid nanoparticles that permeates pig skin was developed. The apparatus includes a Franz diffusion cell with computer-controlled sampling that allows collection of acceptor solution with automatic compensation for sample withdrawing, and a C-18 reversed-phase monolithic column integrated in a typical Flow Injection Analysis (FIA) set-up where separation between caffeine and other matrix elements is performed before spectrophotometric quantification at 273 nm. Several parameters regarding chromatographic analysis (propulsion element, column length, mobile phase composition, and flow rate) were studied along with the establishment of the sampling procedure. Under the selected conditions (monolithic column Chromolith (R) RP-18 15 mm x 4.6 mm i.d., acetonitrile: water 10:90 (v/v), flow rate 0.45 mL min(-1)) a detection limit of 4 mu M and RSD values for caffeine concentration <2% were achieved. High recovery values were obtained when Hepes buffer incubated as acceptor solution in presence of pig skin for 8 h was spiked with caffeine (103 +/- 5%). The developed system also accounts for low organic solvent consumption, low operating costs, low generation of waste and high sample throughput (24 h(-1)). Due to the real time automated sampling and high throughput, transdermal permeation profiles of nanoformulations can be established within a time frame seldom observed by conventional techniques.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 6
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

Recent developments, characteristics and potential applications of screen-printed electrodes in pharmaceutical and biological analysis (2016)
Another Publication in an International Scientific Journal
Couto, RAS; Lima, JLFC; M. Beatriz Q. Garcia
Optical sensors and biosensors based on sol-gel films (2007)
Another Publication in an International Scientific Journal
Paula C A Jeronimo; Alberto N Araujo; Conceicao C B S M Montenegro
Flow injection based methods for fast screening of antioxidant capacity (2009)
Another Publication in an International Scientific Journal
Luis M Magalhaes; Miguel Santos; Marcela A Segundo; Salette Reis; Jose L F C Lima
Extraction and preconcentration techniques for chromatographic determination of chlorophenols in environmental and food samples (2012)
Another Publication in an International Scientific Journal
Paulo de Morais; Teodor Stoichev; Clara C P Basto; Teresa T S D Vasconcelos

See all (193)

Recommend this page Top
Copyright 1996-2025 © Faculdade de Direito da Universidade do Porto  I Terms and Conditions  I Acessibility  I Index A-Z
Page created on: 2025-07-08 at 19:29:30 | Privacy Policy | Personal Data Protection Policy | Whistleblowing