Design and development of a multichannel potentiometer for monitoring an electrode array and its application in flow analysis

Title
Design and development of a multichannel potentiometer for monitoring an electrode array and its application in flow analysis
Type
Article in International Scientific Journal
Year
2002
Authors
Rohwedder, JJR
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Pasquini, C
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Raimundo, IM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Conceicao, M
(Author)
FFUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Montenegro, BSM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Araujo, AN
(Author)
FFUP
View Personal Page You do not have permissions to view the institutional email. Search for Participant Publications View Authenticus page View ORCID page
Couto, CMCM
(Author)
Other
The person does not belong to the institution. The person does not belong to the institution. The person does not belong to the institution. Without AUTHENTICUS Without ORCID
Journal
Vol. 24 No. 4
Pages: 105-110
ISSN: 1463-9246
Scientific classification
FOS: Engineering and technology > Other engineering and technologies
Other information
Authenticus ID: P-000-NV8
Abstract (EN): A versatile potentiometer that works with electrode arrays in flow injection and/or monosegmented flow systems is described. The potentiometer is controlled by a microcomputer that allows individual, sequential multiplexed or random accesses to eight electrodes while employing only one reference electrode. The instrument was demonstrated by monitoring an array of seven flow-through ion-selective electrodes for Ag+ and for three electrodes for Cl-, Ca2+ and K+. The figures of merit of the individual and multiplexed (summed) readings of the electrode array were compared. The absolute standard deviation of the measurements made by summing the potential of two or more electrodes was maintained constant, thus improving the precision of the measurements. This result shows that an attempt to combine the signals of the electrodes to produce a more intense signal in the Hadamard strategy is feasible and accompanied by a proportional improvement in the precision of individual measurements. The preliminary tests suggest that the system can allow for 270 determinations per hour, with a linear range from 1.0 x 10(-2) to 1.0 x 10(-4) mol l(-1) for the three different analytes. Detection limits were estimated as 3.1 x 10(-5), 3.0 x 10(-6) and 1.0 x 10(-5) mol l(-1) for Cl-, Ca2+ and K+, respectively.
Language: English
Type (Professor's evaluation): Scientific
No. of pages: 6
Documents
We could not find any documents associated to the publication.
Related Publications

Of the same journal

Multicommuted flow system for the determination of glucose in animal blood serum exploiting enzymatic reaction and chemiluminescence detection (2003)
Article in International Scientific Journal
Pires, CK; Martelli, PB; Reis, BF; Lima, JLFC; Saraiva, MLMFS
Development of a Multicommutated Flow System with Chemiluminometric Detection for Quantification of Gentamicin in Pharmaceuticals (2010)
Article in International Scientific Journal
Lucia H M L M Santos; Araujo, AN; Boaventura Reis; Montenegro, MCBSM
Design and development of a multichannel potentiometer for monitoring an electrode array and its application in f ow analysis (2002)
Article in International Scientific Journal
Jarbas J. R. Rohwedder; Celio Pasquini; Ivo M. Raimundo Jr; M. Conceição B. S. Montenegro; Alberto N. Araújo; Cristina M. C. M. Couto
An improved leaping detector for flow analysis applied to iron speciation in drugs (2000)